Fluorescent UV lamp is imported for this accelerated aging test chamber
Can simulate harm by sunlight, rainwater and dew
UV weather resistance chamber uses fluorescent UV lamp to simulate the effect of sunlight, condensation moisture is used to simulate dew
The tested material is placed in a loop program that alternates light and moisture at a certain temperature
The material was tested for accelerated weatherability
UV chamber can re-create hazards that occur outdoors for months or years
Types of hazards include: discoloration, chalkiness, powder light, cracking, turbidity, bubbles, brittle, strength, decay, oxidation, etc.
This machine includes spray device
Can simulate ultraviolet, rain, high temperature, high humidity, condensation, darkness in natural environment for this accelerated aging test chamber
By reproducing these conditions, merge into a loop and let it automatically execute the number of completed loops
In this process, the equipment can monitor blackboard temperature and sink temperature automatically
Control the device by configuring radiation measurement (optional), irradiance can be measured and controlled, stabilize the irradiance at 0.76 W/m2/340 nm or a set value, greatly extend the service life of the lamp
Specifications
Working chamber dimensions: 1130 mm × 400 mm × 500 mm
External size: 1300 mm × 1400 mm × 700 mm
Lamp distance to product: adjustable (max 450 mm)
The shortest parallel surface distance between the specimen and the lamp surface: about 100 mm
Wavelength range: UV-A, wavelength range 315 to 400 nm / UV-B, wavelength range 290 to 315 nm
Radiation intensity: 1.2 W/m2/340 nm
Temperature resolution: 0.1 °C
Luminous temperature range: 50 °C to 70 °C / temperature tolerance ±3 °C
Condensation temperature range: 40 °C to 60 °C / temperature tolerance ±3 °C
Blackboard thermometer measuring range: 30 to 80 °C / tolerance ±1 °C
Temperature control type: PID self setting temperature control mode
Humidity range: about 45% to 70% R.H (illumination condition) / 98% (condensation state)
Sink requirements: water depth is not greater than 25 mm and there is an automatic water supply controller
Standard specimen size: 1000 × 400 mm
Recommended environment for instrument use: 5 to 35 °C, 40% to 85% R·H, distance to the wall is 300 mm
Chamber Structure
Spray header
Box door
UV lamp
Fan
Compliance With International Standards
ASTM G 153, ASTM G 154, ASTM D 4329, ASTM D 4799, ASTM D 4587, SAE, J2 020, ISO 4892 all current UV aging test standards
Power Control
Blackboard temperature monitoring: standardized Pt-100 blackboard temperature sensor
Accurate control of the sample surface temperature during the test process
Blackboard temperature setting range: BPT 40 to 75 °C
Actual maximum temperature of protection device inside the machine is 93 °C ± 10%
Blackboard temperature control accuracy: ±0.5 °C
Tank temperature monitoring: one test section is the dark condensation process in the loop test process, it requires the box to be able to produce high temperature saturated water vapor, when the steam hits a relatively cold surface of the sample, dew will condense on the surface of the sample
The tank is located at the bottom of the chamber with built-in electric heater
Tank temperature control range: 40 to 60 °C
The test chamber is equipped with a time controller, range 0 to 530 H, backup data for power fail
Safety protection device
Temperature protection in the chamber: when the temperature in the chamber exceeds 93 °C ± 10%, machine will automatically cut off the supply of the lamp and heater, go into equilibrium stage and cool
Low water tank alarm, prevent heater from burning
Standard Sample Holder
Chamber material: inner material use SUS304# stainless steel plate
Outer shell use SUS304# stainless steel plate
The sample frame is made of stainless steel and aluminum alloy for easy access
Whole Machine Profile
Dimension: about 1600 mm × 1200 mm × 700 mm
Weight: about 150 kg
Mainframe Working Environment
Power requirement: 220 volts and 50/60 Hz
Environment: 5 to 35 °C, 0 to 80% RH, excellent ventilation, indoor environment cleaning
Area of work: about 234 × 353 cm
Draining: a drain ditch is needed on the floor near the main engine
For ease of movement, foot wheel is installed at the bottom of the instrument, position of the test machine is fixed with a u-ring
Control Instrument
The device uses a true color touch screen, PID temperature intelligent control device, high precision temperature control and good stability
Request a Quote
✓
Request sent
Thank you, our team will reply to you shortly.
Direct access to our team
Email American Plant & Equipment
Tell us what you want to buy, source, finance, or sell. A machinery specialist will follow up with you.
✓
Thank you. Your request is with the American Plant and Equipment team.
A machinery specialist will follow up with you.
Machinery updates
Subscribe to our email list
Tell us which industries you follow and we will send you the machines that match.
✓
You are on the list.
We will email you when machines that match your interests come in. Unsubscribe from any message.