REFERENCE NUMBER: A-2492
1000 L STABILITY CHAMBER
SPECIFICATIONS:
INNER CHAMBER SIZE: 1 X 1 X 1 mm
EXTERNAL DIMENSION: 1.2 X 1.97 X 2.17 mm
TEMPERATURE RANGE: 70 TO 180°C
HUMIDITY RANGE: 10% TO 98% RH, WHEN THE TEMPERATURE IS ABOVE 60°C, THE HUMIDITY CAN ACHIEVE 10% RH
TEMPERATURE & HUMIDITY FLUCTUATION: ±0.2℃, ±0.5% RH
INDICATION RESOLUTION: 0.01℃, 0.1% RH
TEMPERATURE & HUMIDITY UNIFORMITY: ±2℃, ±3% RH (TESTED WITH THE EMPTY CHAMBER, 30 MINUTES AFTER STABILIZATION)
HEAT UP AND COOL DOWN TIME: HEATING RATE: AVERAGE 3 ℃/MINUTES; COOLING RATE: AVERAGE 1 ℃/MINUTE (UNLOADED TEST)
MATERIAL: INNER CHAMBER: SUS#304 STAINLESS STEEL EXTERIOR STRUCTURE: STAINLESS STEEL PLATE + POWDER COATED
INSULATION: GLASS WOOL & PUF
PERSPIRATION RESISTANT: K STYLE TUBE
ACCESSORIES
• TEMPERATURE RESISTANT SILICONE GASKET: 1 PC
• CABLE PORT:1 PC; OD: 100 mm
• WINDOW LIGHT: 1 PC
• MULTIPANE VIEW WINDOW: 1 PC
• LOADING SHELVES: 2 PCS
REFRIGERATION & HEATING SYSTEM
• ORIGINAL IMPORTED HIGH EFFICIENCY COMPRESSOR
• AIR COOLING CONDENSER
• FIN – TUBE HEATER
• ORIGINAL IMPORTED SOLENOID VALVES
• INNER SPIRAL K – TYPE REFRIGERANT COPPER TUBE
• U – TYPE FINNED ELECTRO - THERMAL TUBE
• HEATING & COOLING SYSTEMS SEPARATED
• SINGLE STAGE REFRIGERATING SYSTEM
HUMIDIFYING & DEHUMIDIFYING SYSTEM
• ELECTRIC & BYTE TYPE HUMIDIFYING SYSTEM
• STAINLESS STEEL HUMIDIFYING BARREL WITH VIEW WINDOW FOR WATER LEVEL
• ADP DEHUMIDIFYING WAY
• OVER – TEMPERATURE AND OVERFLOW PROTECTION UNITS
• HUMIDIFYING & DEHUMIDIFYING SYSTEMS SEPARATED
CIRCLING & DRAIN SYSTEM
• CENTRIFUGAL FAN & IMPORTED MOTOR
• REVERSE UNIFLOW (AIR FLOW DIRECTION)
• FLOW – THROWS WIND
• DRAIN SYSTEM USES SWIRL TYPE OUTLET AND U – TRAP DEPOSITION UNIT
PROGRAMABLE MICRO COMPUTER LCD TOUCH CONTROLLER
• PROGRAMMABLE TOUCH CONTROLLER, USER - FRIENDLY, STABLE AND EASY USE
• PROGRAMS: 99 SEGMENTS PER PROGRAM, 120 PROGRAMS, 120 CYCLING PER PROGRAM
• PID & SSR CONTROL WAY, STABLE AND EASY USE
• HAVE SLOP CONTROL FUNCTION FOR HEATING & COOLING, LINKED PTNS CAN BE EXECUTED
• PID AUTOMATIC CALCULATION
• MICRO COMPUTER MULTI FUNCTION CONTROLLER, MAIN COMPONENTS ’DATA AUTOMATICALLY OUTPUT INCLUDING COMPRESSORS, SOLENOIDS, HEATING UNITS, ETC
• ERRORS OF MAIN COMPONENTS APPEAR ON THE LCD PANEL
• THE BRIGHTNESS OF THE LCD PANEL CAN BE ADJUSTED
• TIME SEQUENCE CONTROL INTERFACE RELAYS HAVE 3 GROUPS WHICH MANAGE THE OUTPUT OF OUTSIDE LOGIC UNITS INCLUDING SIGNAL, MECHANICAL AND POWER, ETC
• PROGRAMS AUTO – SAVE FUNCTION IF POWER CUT OFF SUDDENLY
• HAVE HOLD & ADV BUTTONS INPROGRAMS
• AUTOMATICALLY SET THE HIGH & LOW TEMPERATURE LIMITS AFTER PROGRAMS SET
• PASSWORD SET FUNCTION
• SD CARD SLOT, 2 GB SD CARD, RS232 OR ETHERNET (RJ45)
OPERATING ENVIRONMENT
• AMBIENT TEMPERATURE: +25℃ (2) RH ≤85%
PROTECTION
• ELECTRICITY LEAK PROTECTION
• LIGHTNING STRIKE PROTECTION, ZRN LIGHTNING ARRESTER INSIDE THE CONTROLLER
• SENSOR DATES TRANSFERRED INSIDE THE MICRO COMPUTER
• TEMP. SENSORS TO OPEN CIRCUIT OR SHORT CIRCUIT (+ - OVER)
• THE FIRST LEVEL OVER TEMPERATURE PROTECTION
• ERROR DETECTED UNITS
• THE SECOND LEVEL OVER TEMPERATURE PROTECTION TO ELECTRONIC OVER - TEMP PROTECTION UNIT
• THE THIRD LEVEL OVER - TEMPERATURE PROTECTION
• COMPRESSOR PROTECTION TO REFRIGERANTS PRESSURE PROTECTION & OVERLOAD PROTECTION.
• EXTENDED PROTECTION UNIT TO TWO SUPPLETIVE PORTS
• ERROR PROTECTION TO POWER CUT OFF, ERROR REPORTS SHOWN IN LCD PANEL AND AUDIBLE ALARM STARTS
• DYNAMIC HIGH AND LOW TEMPERATURE PROTECTION
• AMBIENT TEMPERATURE PROTECTION TO THE INNER CHAMBER AND SPECIMENS AUTOMATICALLY RETURN BACK TO AMBIENT TEMPERATURE AFTER THE OPERATION STOP SAFETY PROTECTION
• NON-FUSE PROTECTION SWITCH
• OVERHEAT ALARM PROTECTION
• COMPRESSOR OVERLOAD PROTECTION
• PHASE SEQUENCE PROTECTION
• MOTOR OVER TO CURRENT PROTECTION
Learn More