NovaStar, FE-28, Bacterial Filtration Efficiency Tester, 220V, 1kW, 28.3 L/min (New)

Description

  1. Overview

    1. Safe working environment – uses professional negative pressure biological cabinet to ensure the safety of operators

    2. High negative pressure working cabin: two-stage high efficiency filter, 100% safe discharge

    3. Use two-way six-level Anderson sampling

    4. Built-in peristaltic pump with adjustable flow rate

    5. Special microbial aerosol generator, adjustable spray flow of bacterial liquid, good atomization effect

    6. Industrial grade 10.4-inch color touch screen control, making the operation simpler

    7. Microcomputer is used for control and data processing, which can be seamlessly connected to the laboratory information management system

    8. USB interface, support data transfer

    9. The safety cabinet has built-in LED lighting for easy observation

    10. Built-in ultraviolet UV disinfection lamp

    11. The front-opening sealed glass door is convenient for operation and observation

    12. Use imported isolation operation gloves – safe and reliable

    13. The system has functions of user authority management and audit trail

  2. Specifications

    1. Dimension: 1250 x 700 x 2000 mm

    2. Voltage: 220 volts and 50/60 Hz

    3. Total kW: 1 kW

    4. Floorspace required: 2 square meters

    5. Anode quality control sampling flow: 28.3 L/min, resolution 0.1 L/min, accuracy ±2%

    6. Sampling flow: 28.3 L/min, resolution 0.1 L/min, accuracy ±2%

    7. Spray flow: 0.1 to 10 L/min, resolution 0.1 L/min, accuracy ±5%

    8. Peristaltic pump flow: 0.01 to 3.0 mL/min, resolution 0.01 mL/min, accuracy ±2%

    9. Operating temperature: 0 to 50 °C

    10. Biosafety zone negative pressure: -90 to -120 Pa

    11. Data storage capacity: 20 G

  3. Lead Time

    1. 5 to 7 weeks from receipt of order