NovaStar, EC102, Constant Temperature and Humidity Test Chamber, 150L, -20 to +150C, 20% to 98% RH
Scroll to zoom
100%
Description
Product Usage
Simulate product in the condition of climatic environment (high and low temperature operation and storage, temperature loop, high temperature and humidity, low temperature and humidity, moisture condensation etc), inspect product adaptive capacity and character if change.
Conform to international standard (IEC, JIS, GB, MIL) reach international measurement program consistency (including test process, condition and method).
Limited Sample
Applicable for below: electrician, electronics, components and parts; material storage and transportation under the condition of high and low temperature; compatibility test when use.
Prohibition for below: inflammable, explosive, effumability substance test or storage; corrosive substances test and storage; biological sample test and storage; strong electromagnetic emitter sample test and storage.
Volume, Dimension And Weight
Marked volume: 150 L.
Inner box dimension: 500 x 600 x 500 mm.
Carton dimension: 1050 x 1750 x 1080 mm.
Weight (about): 220 kg.
Structure Performance
Test environment condition: environment temperature +25 °C, relative humidity ≤ 85% test chamber without sample.
Test method: GB/T 5170.2-1996 temperature testing equipment, GB/T 5170.5-1996 humid heat test equipment (only damp-heat type).
Temperature range: -20 °C to +150 °C.
Temperature fluctuation: ±0.5 °C.
Temperature distribution precision: ±2.0 °C.
Humidity temperature range: 20% to 98% R.H. (only damp-heat type).
Humidity distribution precision: ±3.0% (only damp-heat type).
Heating up time: -20 °C to +150 °C heating up 4.0 °C/min.
Cooling time: +150 °C to -20 °C cooling 1.0 °C/min.
Load condition: test space not exceed 3/4 of total space, no exist heat emission.
Product testing frame: double-deck test frame, each layer of bearing 50 kg, available to adjust height.
Structural Feature
Heat preservation structure: outer wall material high quality stainless steel 304, cold rolled steel plate spray paint; inner wall material high quality stainless steel 304; camber heat preservation material flame retardant fire resistant rigid polyurethane foaming heat preservation material.
Air adjustment channel: centrifugar fan, air return method blowing wind above equipment, return wind below; heater, evaporator, outfall, temperature sensor.
Test chamber standard configuration: window 250 x 320 x 40 mm 3 layer vacuum; anti-explosion grip; hinge SUS #304; chamber inner light PL light 11 W; fairlead φ50 mm 1 pcs.
Door: single door, left open, grip on the right, with observation window 250 x 320 mm floodlight, window frame, door frame anti-condensation electrothermal device.
Control panel: controller display screen, over-temperature protection settings.
Machine room: refrigerating unit, water pan, drain hole.
Power distribution control cabinet: distribution board, exhaust fan.
Heater: heating system fin heat sink tube stainless steel electric heater air loop method.
Humidifier: boiler humidifying system, the return loop of the air duct system.
The power cord hole and drain hole: located on the back of chamber.