NovaStar, Respirator Resistance Tester, 0-100 L/min, 220V 50Hz

Description

  1. Overview
    1. Suitable for medical and daily protective masks to test whether the resistance value of exhalation and inhalation to human body is up to the standard
    2. Relative standards: GB2626-2019, GB/T32610-2016, TAJ1001-2015 and so on
  2. Specifications
    1. Power: 220 V, 50 Hz
    2. Flowmeter range: 0 to 100 L/min, precision: 3%
    3. Micropressure transducer: ±1000 Pa, resolution 1 Pa
    4. VC defaults to (85 ± 1) L/min (can be set according to the demands)
    5. Specifications of test head die can be changed as required (standard GB2626, 3 pieces)
    6. Automatic test and evaluation of test results; eligibility criteria: expiratory resistance less than 250 Pa, inspiratory resistance less than 350 Pa
    7. Power: AC 220 V ± 10%, 50 Hz / 220 V, 60 Hz
    8. Instrument size: 620 × 650 × 850 mm (L × W × H)
    9. Weight: about 30 kg