IDM Instruments, 1737 Series, Oxygen & Carbon Dioxide Analyzer, Bluetooth Data Uploads, 3700 Measurement Memory (New)

Description

  1. Overview
    1. Accurate and drift-free measurement of head space for closed packages and cans
    2. Automatically zeroed whenever the sample is ambient air and only requires calibration gas for the ppm level sensor
    3. Carbon dioxide is also zeroed whenever the sample is ambient air, and uses a simple keyboard operation for the span calibration
    4. Sample gas flow rate is measured using a true mass flow sensor and allows the unit to maintain a constant flow rate and alarm if there is a blockage
    5. Operator does not need to touch the analyzer once the product has been selected
    6. Each time the sample gas returns to ambient air, the alarms are checked, and the min/max measurements are stored and the display is updated
  2. Features
    1. Bluetooth data uploads
    2. 200 product categories
    3. Power or battery operation
    4. Sample gas flow measurement and control
    5. Large graphic display
    6. Performs automatic zero calibration when reading ambient air
    7. Data memory for 3,700 measurements
  3. Applications
    1. Flexible packaging
    2. Food packaging
    3. Pharmaceuticals
  4. Available Models
    1. Model 1737-1: Oxygen measurement only (0.1 to 25%) MAP packaging
    2. Model 1737-2C: Oxygen and CO2 measurement (30ppm to 25%) and carbon dioxide MAP packaging
    3. Model 1737-3C: Oxygen and CO2 measurement (0.1 to 96%) and carbon dioxide MAP packaging
  5. Specifications
    1. Measuring range: Oxygen 0.1 to 25%, 30ppm to 25%, or 0.1 to 96%; CO2 0 to 100%
    2. Accuracy: Oxygen (25-96%) ±2%; Oxygen (10-25%) ±0.5%; Oxygen (0.1-10%) ±0.01%; CO2 (0-40%) ±1.5%; CO2 (40-95%) ±2.0%
    3. Resolution: Oxygen 30 to 96% | 0.1%; Oxygen 1.00 to 29.99% | 0.01%; Oxygen 0.001 to 0.999 | 0.001%; Oxygen 100 to 999ppm | 1ppm; Oxygen 0.1 to 99.9ppm | 0.1ppm; CO2 0.1 to 100% 0.1%
    4. Head space volume: Oxygen sample flow 50cc/m: 1% to 10% requires 15cc, 100 to 1,000ppm requires 40cc; Oxygen sample flow 150cc/m: 1% to 10% requires 30cc, 100 to 1,000ppm requires 90cc; CO2 sample flow 150cc/m: 1% to 40% requires 20cc
    5. Warm up time: 1 minute
    6. Gas sample flow range: Automatic control 50 to 250cc/min; Manual control 30 to 350cc/min
    7. Gas connection: 1/8” Swagelok tube connection
    8. Communications: Data transfer via Bluetooth to PC
    9. Power supply: 12V or AA batteries (x7)
    10. Dimensions: L 280 x W 180 x H 115mm
    11. Weight: Analyzer 2kg; Power pack 0.5kg