NovaStar, EC136, Salt Spray Corrosion Test Chamber, 108 to 1440 Liters, AC220V (New)

Description

  1. Overview
    1. Salt spray corrosion test chamber suitable for neutral salt spray test (NSS).
    2. Heating method: test room adopts steam direct heating method, temperature rise fast and uniform, reduce the standby time; the heating tube is made from titanium tube with high corrosion resistance.
    3. Laboratory commodity shelf uses plane dividing shelf; commodity shelf is the 10 mm toughened glass rod with corrosion resistance.
    4. Plastic steel V type shelf and glass rod placing objects make sure the sample 30 degree bevel angle; the angle can be arbitrarily adjusted; the direction of spray falling around is the same as the direction of spray receiving; test piece is set with large quantity.
    5. Sample supplement jar adopts hidden and water level perspective window attached; it is easy to clean and not easy cracking.
    6. Salt spray test aims at determining the product corrosion resistance after various kinds of materials surface suffering electroplating, anodizing, painting, rust-proof oil corrosion resistance treatment.
  2. Key Specifications
    1. Inside chamber dimension (W x D x H mm): 600 x 450 x 400, 900 x 600 x 500, 1200 x 1000 x 500, 1600 x 1000 x 500, 2000 x 1200 x 600
    2. Outside chamber dimension (W x D x H mm): 1070 x 600 x 1180, 1410 x 880 x 1280, 1900 x 1300 x 1400, 2300 x 1400 x 1400, 2700 x 1500 x 1500
    3. Test chamber volume: 108 liters, 270 liters, 600 liters, 800 liters, 1440 liters
    4. Brine tank: 15 liters, 25 liters, 40 liters
    5. Control method: LED meter
    6. Plastic plate thickness: 5 mm
    7. Temperature of test chamber: brine test method 35 °C ± 1 °C
    8. Temperature of pressure barrel: brine test method 47 °C ± 1 °C
    9. Brine temperature: 35 °C to 50 °C
    10. Brine concentration: 5% sodium chloride solution concentration
    11. Compressed air pressure: 1.00 ± 0.01 kgf/cm²
    12. Spray volume: 1.0 to 2.0 ml/80 cm²/hour (gather at least 16 hours, take its average)
    13. Relative temperature of test chamber: above 85%
    14. pH-value: neutral 6.5 to 7.2
    15. Spray method: continuous spray or intermittent spray
    16. Power supply: AC220V 1Φ 10A, AC220V 1Φ 15A, AC220V 1Φ 30A
  3. Temperature Range
    1. Laboratory temperature: 35 °C to 50 °C (can be arbitrarily set)
    2. Pressure barrel temperature: 47 °C to 63 °C (can be arbitrarily set)
  4. Temperature Rising Speed
    1. Laboratory room temperature → 35 °C: about 35 minutes
    2. Pressure barrel room temperature → 47 °C: about 15 minutes
    3. Temperature accuracy: ± 1 °C
  5. Specifications
    1. Test chamber temperature: 35 to 50 degrees C