Map Industries, Ethylene Oxide Sterilization System, 24 m³ Capacity, 1820x1820x7250mm (New)

Description

  1. Overview
    1. Ethylene oxide sterilization system with double-door chamber
    2. Chamber is constructed of stainless steel with a stainless steel heating jacket
    3. Designed to withstand pressure equal to 20 psig; entire cycle is performed in vacuum
    4. Doors are manually operated with a radial arm design and a special locking mechanism that prevents opening after a certain pressure and vacuum
    5. Back plate is specially made to withstand the rigors of pressure and vacuum
  2. Chamber Specifications
    1. Chamber dimensions: 1820 mm wide x 1820 mm high x 7250 mm long
    2. Double door configuration: one front door and one back door
    3. Door gasket is a single piece design made of molded silicon rubber for excellent sealing and long durability
    4. Sufficient number of chamber openings provided for mounting valves and sensors
    5. Welds are free of pits and crevices; internal welds are ground smooth and flush with interior chamber walls
    6. Chamber is insulated and sheathed with SS 304 for a pleasing appearance and excellent thermal barrier to reduce heat load
    7. Sheathed insulation is impact resistant to prevent damage and allow easy maintenance
  3. Heating System
    1. Hot water circulation system through piping around the chamber
    2. Separate tank with adequate storage capacity and electric heaters
    3. Heated water is circulated throughout the chamber periphery with circulation pumps
  4. Mechanical Equipment and Valves
    1. Sturdy and reliable mechanical components used in assembly, including valves, pumps, and regulators
    2. Components chosen from reputed suppliers to simplify calibration, maintenance, replacement, and servicing
  5. Auxiliary Equipment
    1. Vacuum pump suction system to evacuate chamber to 22. Hg; system cycles empty chamber in less than 30 minutes; vacuum cycle is controlled automatically
    2. Filtered air inlet/break assembly with filter housing and removable filter cartridge for vacuum break at end of vacuum pulses
    3. Specially designed gas evaporator ensures gas enters chamber in vapor state only; gas is conditioned automatically to predetermined temperature by maintaining vaporizer temperature at a preset level
    4. Steam generator for humidification; humidity stage is carried out prior to introduction of ethylene oxide and CO2 mixture
  6. Automation and Electronic Control System
    1. Dedicated automatic controller controls entire sterilization cycle; all parameters can be set and are monitored continuously
    2. Control system monitors utilities and informs operator if any utilities are missing or needed per chosen cycle requirement
    3. Loss of any monitored parameter during cycle generates audible and visual alarm
    4. Standard features include program selection mode for multiple cycle programs, multiple password security system, setting of various parameters to alter cycle with process control limits, in-built aeration facility, safety from overpressure and leakage, and touch screen controls
    5. Setup menu allows operator to select from pre-programmed cycles; each cycle consists of segments or phases; once programmed, same cycle repeats until modifications are made
    6. Controller advances system through required phases as conditions for each portion are completed; operator display shows current process values such as chamber temperature and pressure; all parameter values are printed
    7. Provision to change over from auto to manual operation in case of unlikely failure of automation system
    8. Separate chart recording system ensures batch record can be maintained even if auto system fails and cycle is operated in manual mode
  7. Safety Features
    1. Temperature sensor and electronic pressure sensor linked to automation ensure reliability and failure-safe operation
    2. Constant watch over entries and breaks in diagnostics from automation warns operator via messages on indicator and buzzer
    3. Physical safety measures include leak protection, over-pressure protection, locked door sensing, and pressure lock for door
    4. Operator protection includes aeration for reduction in residual exposure, guidance for air exhaust system in chamber and allied rooms, safety instructions, and fire safety instructions
    5. Safety alarm in case of any malfunction, errors, or utility failures
  8. Additional Features
    1. Evacuation of chamber during aeration is programmable; auto-manual changeover possible in case of failure of auto system; password protection for change of instructions
    2. After closing chamber door, entire sterilization operation (evacuation, feeding of gas, degassing, and fresh air inlet) is carried out automatically
    3. Sterilizer controlled using dedicated automation controller that controls entire automatic cycle; all parameters like pressure, vacuum, temperature, and humidity are monitored
    4. Control system monitors plant utilities and equipment functions; loss of any essential utility during cycle generates audible and visual alarm
    5. Multiple program cycle for different applications as required; password security system
    6. Setting of various parameters to alter cycles; safety from overpressure and leakage
    7. Operator starts cycle and controller advances system through required phases as conditions for each portion are completed; cycle advances through necessary segments as conditions are achieved
    8. During cycle, operator display shows current process values such as chamber temperature and pressure along with other set points; all parameter values are recorded on a data logger
    9. Provision to take complete cycle in manual mode in case of failure in auto mode; if auto system fails, sterilizer can be run in manual mode
  9. Specifications
    1. Capacity: 24 cubic meters