NovaStar, SZ133, Ethylene Oxide Sterilizer, 60 kVA, 1120x120x150cm Chamber, Automatic Door Cabinet (New)

Description

  1. Overview
    1. Product form: automatic door cabinet
    2. Heating mode: electrical heating
    3. Power supply: AC380V, 50 Hz (220V available)
    4. Power: 60 KVA
  2. Specifications
    1. Sterilizing chamber size (L x W x H): 1120 x 120 x 150 cm
    2. Outer box size (L x W x H): 1170 x 190 x 226 cm
    3. Sterilizing chamber material: 304 anticorrosive stainless steel
    4. Outer box material: stainless steel (1.2 mm thick)
    5. Sterilization temperature: 50°C ±3°C
    6. Working pressure: -75KP to 50 KPA
    7. Maximum pressure resistance: -80KPA
    8. Sterilization humidity: 45%-75%
    9. Selected gas: mixed gas (80% EO, 20% carbon dioxide)
    10. Ethylene oxide sterilization concentration: 0.8 to 1.2 g/L
    11. Sterilization time: 0-99 hour adjustable
    12. Vacuum speed: 25 minutes to -50KPA
    13. Air replacement: automatic ventilation, air exchange frequency and interval can be set freely
  3. Control And Display
    1. Automatic control system: one-click start to end; the whole process is automatically completed with print record
    2. Display status: displays temperature, humidity, pressure, sterilization time, residual gas treatment time, number of times
    3. Printer: the whole process of sterilization information record printing, in order to verify the check
  4. Sensors And Systems
    1. Humidity sensor: the built-in humidity sensor automatically detects the humidity in the cavity to ensure that the humidity in the cavity is not lower than 45% RH
    2. Pressure sensor: the inner pressure sensor automatically detects the pressure inside the cavity to ensure the accuracy and efficiency of the pressure
    3. Temperature sensor: the built-in temperature sensor automatically detects the cavity temperature to ensure that the cavity temperature remains within the accurate range of 50°C ±3°C
    4. Humidifying system: pulse humidification in the vacuum state ensures that the humidity in the cavity is uniform and controllable, and the humidity of the sterilizing chamber should be controlled within the range of 45-80% relative humidity
    5. Residual gas treatment unit: filtration efficiency of filter for particle diameter ≥ 0.2mm, not less than 99.5%
    6. Sterilization chamber leakage rate: the leakage rate was measured under the condition of no load, the temperature of sterilizing chamber was constant, the test pressure was -50 KPA, the measuring time was 1 h, and the average leakage rate was less than 0.1 KPA / min
    7. Sealing door: hydraulic automatic opening and closing, easy operation, good sealing
    8. Vacuum rate: under the condition of no load, the time of vacuum pumping from normal pressure to -60 KPA should be less than 30 minutes
    9. Vacuum system: oil-free vacuum pump, simple and convenient, rapid vacuum rate, long service life
    10. Heating system: incorporate explosion-proof electric heating device, more convenient and quicker
  5. Additional Features
    1. Tray rack: 304 stainless steel tray rack; trolley can be selected
    2. Multiple protective functions: it has the function of overvoltage and overtemperature protection, which is safer and more reliable
    3. Alarm system: high temperature alarm system, ultra-low pressure alarm system