NovaStar, SZ134, 26m3 Cabinet Volume, Aeration Chambers (New)

Description

  1. Overview
    1. Cabinet volume: 26 m³
    2. Inside cabinet size (length x width x height): 4300 x 2800 x 2200 mm
    3. Overall outer size (length x width x height): 4800 x 2200 x 2700 mm
    4. Size can be adjusted according to customer requirement.
    5. Pressure range chamber supported: designed range -95 kPa to +80 kPa, suggested pressure -90 kPa to +60 kPa.
    6. Product loading way: suggest that the chamber is set in a pit to make the bottom of the internal chamber and the floor even; device provided to connect chamber bottom to ground; suggest pallet loading way.
  2. Body Construction
    1. Cabinet: stainless steel SUS304, Δ = 8 mm.
    2. Water jacket: stainless steel #304, Δ = 6 mm.
    3. Housing: stainless steel.
    4. Foot: U-bar steel 14# / Q235.
    5. Foot support: steel sheet 10 / Q235.
    6. Insulating layer: ceramic fiber.
  3. Door Construction
    1. Door: stainless steel SUS304, Δ = 8 mm.
    2. Door frame: rectangular steel 120 x 60 / Q235.
    3. Door water jacket: stainless steel.
    4. Door seal flange: stainless steel.
    5. Door seal way: air internal feeding.
    6. Housing: stainless steel.
    7. Insulation layer: ceramic fiber.
    8. Pneumatic revolving door (two), door sealing way: air inject inside.
  4. System Functions and Properties
    1. System design standard: EN1422, ISO11135-2014.
    2. System control mode: by switching, soft manual and full-automatic.
    3. Open/close door way: pneumatic revolving door.
    4. Door quantity: two.
    5. Data record way: interval mode, warning triggered mode.
    6. Historical curve record way: continuous time interval mode.
    7. Data save time: 10 years (if hard drive is not damaged).
    8. Temperature sensor precision: display range 0 to 100 °C, resolution 0.1 °C.
    9. Temperature control precision: cabinet wall temperature difference within ±5 °C.
    10. Cabinet inside temperature difference is within ±3 °C.
    11. Pressure sensor precision: display range -100 to +150 kPa, resolution 0.1 kPa.
    12. Vacuum control mode: continuous mode, stepping mode, variable frequency regulating mode or proportional valve regulating mode.
  5. Specifications
    1. Voltage: 480V